We are a highly renowned firm engaged in the provision of highly qualitative Anesthesia Delivery System. These systems are manufactured in compliance with industrial quality standards. These are equipped with anesthesia ventilator which operates on different ventilation modes. These are also equipped with a gas supply pipeline and a mechanical pressure gauge. The automatic bypass changes soda lime during surgery. The vent switch permits manual ventilation and mechanical detection. Alarm settings are present for negative pressure, mains failure, supply failure, exhausted battery and low capacity battery. (Make USA) |
|
Technical Specifications : |
12151 62nd St N #5 |
Largo FL 33773 USA |
Phone: (1) 727-531-8434 |
Fax: (1) 727-531-8436 |
Anesthesia Ventilator |
Ventilation Mode VCV, PCV, SIMV+VCV+PSV, SIMV+PCV+PSV, Manual, Standby |
Tidal Volume (VT) 20~1500 mL |
Frequency (f) 1~60 bpm |
I :E 3:1~1:6 (Increment: 0.5) |
Ti 0.1~10 s (Increment: 0.1 s) |
I nspiratory Pause (Tp) OFF, 5%~50% |
E-PEEP I ntegrated electronic PEEP OFF, 4~30 cmH2O |
Flow Trigger (FTRIG) 1~15 L/min |
Pressure Control 5~70 cmH2O |
Pressure Support 5~60 cmH2O |
Pressure Limit PEEP+5~PEEP+70 cmH2O |
Monitors C, R, VTE, VTI, MV, MVspn, f, fspn, I:E, Ppeak, |
Pmean, Pplat, Pmini |
Pneumatics |
Gas Supply Pipeline: O2, N2O and Air; (optional) Yoke: 1x O2; 1x N2O; |
Pressure Gauge Mechanical |
Flowmeter Electronic for O2, N2O and Air; O2 and N2O linkage and N2 cutoff function |
Bellows Volume Adult: 0~1500 mL, pediatric (optional), 0~300 mL |
APL Valve 2~70 cmH2O |
CO2 Absorber Volume: 1.5 L (1.35 Kg) |
By-Pass Automatic; change soda lime during surgery. |
Bag/Vent Switch S witch for manual ventilation and mechanical ventilation |
|
Monitoring: |
Omni IIINI BP, SPO2, ECG, Temp, IBP, CO, Etco2, Anesthetic agent (optional) |
Waveform P-T, F-T, V-T, SpO2-T (optional), CO2-T (optional) |
Loop V-P, F-V, F-P (optional) |
|
Alarm Setting: |
VT, MV, f, FiO2, Airway pressure, High continuous airway pressure, |
Negative pressure, Apnea, O2, Supply failure, Mains failure, Battery low capacity, Battery exhausted |
|
Alarm Mute: |
< 120s |
|
Vaporizer: |
2 stations selectatec mount |
|
Physical: |
S creen 10.4 inch TFT screen (optional touch screen) |
Dimension 1570× 850× 740 mm3 (H×W× D) |
N et Weight 86 Kg (Kg only framework) |
Power 100 to 240 VAC, 50/60 Hz, 10A (Max.) |
Lithium Battery 120 minutes |
Communication VGA, RS232, RJ45, HL7 (with Omni III patient montior) |
|
|